SF6 P1:p MANUAL Ion Science Ltd
38 of 41
Unrivalled Detection. www.ionscience.com
Error messages
Fig. 1
ERR19 (Excess Sensor
Current)
- Please replace Sensor!
The current drawn by the SmartSensor is too high and cannot be controlled
to the setpoint. Possibly liquid in the SmartSensor, or handset electronics is
defective.
REMEDY: Replace the SmartSensor. Replace the handset electronics.
>>>> ALLGEMEIN
WARNING:
SmartSensor lifetime is
nearly over!
Please use new P1:P
SmartSensor!
The SmartSensor has been operated for over 300 operating hours and is
approaching the end of its service life.
REMEDY: Replace the SmartSensor as soon as possible.
NOT READY !
SmartSensor lifetime is
over!
Please use new P1:P
SmartSensor!
The SmartSensor has reached its service life limit of 320 hours and must no
longer be used.
REMEDY: Replace the SmartSensor.