GASCHECK TESLA MANUAL Ion Science Ltd
Page 13 of 25
Unrivalled Detection. www.ionscience.com
Functions
Gas selection
GasCheck Tesla is calibrated using Helium gas at manufacture, however other
gases can be detected and measured using a gas table within the instrument.
Although GasCheck Tesla can't differentiate between gases it can quantify other
gases using response factors stored in its internal gas table. Gas groups 0, 1 2, 3,
and 4 offer a generic sensitivity.
The following list indicates response factors for specific gases however for advice on
gases not listed please contact Ion Science Ltd for advice.
Minimum sensitivity
cc/sec (ml/sec)
Minimum sensitivity
cc/sec (ml/sec)
To select an alternative gas use the 'Up and Down' keys to select the desired gas
and then press the 'ENTER' key to confirm the selection, at which point a tick will
appear. Press the 'ENTER or Esc' key to exit into the 'Main menu'.
If the gas to be detected does not appear in the instruments internal gas table then
one of the five 'Gas Groups' can be selected. 'Gas Groups' have a variation of
sensitivities. (Group 1 being high sensitivity through to Group 5 being low sensitivity)
Contact Ion Science Ltd for advice. Gas groups allow the instrument to calculate
leak rates with reasonable accuracy.