GASCHECK 3000 & 3000IS MANUAL Ion Science Ltd
Page 24 of 28
Unrivalled Detection. www.ionscience.com
Maintenance
Fault finding
Try another battery set (safely)
Remove & replace battery clip (safely)
Replace batteries (safely)
Back light drains battery life
Excessive use of backlight
Avoid or reduce use of backlight
Software locked out by hardware
Remove & replace batteries (safely)
Display appears blank/dim
Contrast control setting wrong
Turn off & on; press & hold contrast
Press zero or change from Peak hold
Range not set in Auto-range
Press & hold Range until Auto-range
He range will not show –5
Unit with nozzle ON and/or range
Reset via Nozzle and Range keys
Gas Group not set correctly
Press & hold Gases until shown OK
Nozzle/Peak is not set correctly
Press & hold Nozzle until both correct
Motor failed & probe not blocked
Return to Ion Science for repair
Instrument readings noisy
Instrument readings noisy
Test environment contaminated
Move item under test into cleaner area
3000is unit will not turn on
Return to supplier for repair